25 Physical Methods to Control Microorganisms
In this section
Content in this section includes:
For thousands of years, humans have used various physical methods of microbial control for food preservation. Common control methods include the application of high temperatures, radiation, filtration and desiccation (drying), among others. Many of these methods non-specifically kill cells by disrupting membranes, changing membrane permeability or damaging proteins and nucleic acids by denaturation, degradation or chemical modification.
Heat
Heating is one of the most common—and oldest—forms of microbial control. It is used in simple techniques like cooking and canning. Heat can kill microbes by altering their membranes and denaturing proteins. The thermal death point (TDP) of a microorganism is the lowest temperature at which all microbes are killed in a 10-minute exposure. Different microorganisms will respond differently to high temperatures, with some (eg endospore-formers such as C. botulinum) being more heat tolerant. A similar parameter, the thermal death time (TDT), is the length of time needed to kill all microorganisms in a sample at a given temperature. These parameters are often used to describe sterilisation procedures that use high heat, such as autoclaving. Boiling is one of the oldest methods of moist-heat control of microbes and it is typically quite effective at killing vegetative cells and some viruses. However, boiling is less effective at killing endospores as some endospores are able to survive up to 20 hours of boiling. Additionally, boiling may be less effective at higher altitudes, where the boiling point of water is lower and the boiling time needed to kill microbes is therefore longer. For these reasons, boiling is not considered a useful sterilisation technique in the laboratory or clinical setting.
Many different heating protocols can be used for sterilisation in the laboratory or clinic, and these protocols can be broken down into two main categories: dry-heat sterilisation and moist-heat sterilisation. Aseptic technique in the laboratory typically involves some dry-heat sterilisation protocols using direct application of high heat, such as sterilising inoculating loops (Figure 7.2). Incineration at very high temperatures destroys all microorganisms. Dry heat can also be applied for relatively long periods of time (at least 2 hours) at temperatures up to 170 °C by using a dry-heat steriliser such as an oven. Moist-heat sterilisation is typically the more effective protocol because it penetrates cells better than dry heat.

Autoclaves
Autoclaves rely on moist-heat sterilisation. They are used to raise temperatures above the boiling point of water to sterilise items such as surgical equipment from vegetative cells, viruses and especially endospores, which are known to survive boiling temperatures, without damaging the items. Charles Chamberland (1851–1908) designed the modern autoclave in 1879 while working in the laboratory of Louis Pasteur. The autoclave is still considered the most effective method of sterilisation (Figure 7.3). Outside laboratory and clinical settings, large industrial autoclaves called retorts allow for moist-heat sterilisation on a large scale.
In general, the air in the chamber of an autoclave is removed and replaced with increasing amounts of steam trapped within the enclosed chamber, resulting in increased interior pressure and temperatures above the boiling point of water. The two main types of autoclaves differ in the way that air is removed from the chamber. In gravity displacement autoclaves, steam is introduced into the chamber from the top or sides. Air, which is heavier than steam, sinks to the bottom of the chamber, where it is forced out through a vent. Complete displacement of air is difficult, especially in larger loads, so longer cycles may be required for such loads. In pre-vacuum sterilisers, air is removed completely using a high-speed vacuum before introducing steam into the chamber. Because air is more completely eliminated, the steam can more easily penetrate wrapped items. Many autoclaves are capable of both gravity and pre-vacuum cycles, using the former for the decontamination of waste and sterilisation of media and unwrapped glassware and the latter for sterilisation of packaged instruments.

Standard operating temperatures for autoclaves are 121 – 132 °C , typically at a pressure of greater or equal to 205 kPa or 2.02 atm (about double atmospheric pressure). The length of exposure depends on the volume and nature of material being sterilised, but it is typically 20 minutes or more, with larger volumes requiring longer exposure times to ensure sufficient heat transfer to the materials being sterilised. The steam must directly contact the liquids or dry materials being sterilised, so containers are left loosely closed and instruments are loosely wrapped in paper or foil. The key to autoclaving is that the temperature must be high enough to kill endospores to achieve complete sterilisation.
Because sterilisation is so important to safe medical and laboratory protocols, quality control is essential. Autoclaves may be equipped with recorders to document the pressures and temperatures achieved during each run. Additionally, internal indicators of various types should be autoclaved along with the materials to be sterilised to ensure that the proper sterilisation temperature has been reached (Figure 7.4). One common type of indicator is the use of heat-sensitive autoclave tape, which has white stripes that turn black when the appropriate temperature is achieved during a successful autoclave run. This type of indicator is relatively inexpensive and can be used during every run. However, autoclave tape provides no indication of length of exposure, so it cannot be used as an indicator of sterility. Another type of indicator, a biological indicator endospore test, uses either a strip of paper or a liquid suspension of the endospores of Geobacillus stearothermophilus to determine whether the endospores are killed by the process. The endospores of the obligate thermophilic bacterium G. stearothermophilus are the gold standard used for this purpose because of their extreme heat resistance. Biological endospore indicators can also be used to test the effectiveness of other sterilisation protocols, including ethylene oxide, dry heat, formaldehyde, gamma radiation and hydrogen peroxide plasma sterilisation using either G. stearothermophilus, Bacillus atrophaeus, B. subtilis, or B. pumilus spores. In the case of validating autoclave function, the endospores are incubated after autoclaving to ensure no viable endospores remain. Bacterial growth subsequent to endospore germination can be monitored by biological indicator endospore tests that detect acid metabolites or fluorescence produced by enzymes derived from viable G. stearothermophilus. A third type of autoclave indicator is the Diack tube, a glass ampule containing a temperature-sensitive pellet that melts at the proper sterilisation temperature. Endospore strips or Diack tubes are used periodically to ensure the autoclave is functioning properly.

Pasteurisation
Although complete sterilisation is ideal for many medical applications, it is not always practical for other applications and may also alter the quality of the product. Boiling and autoclaving are not ideal ways to control microbial growth in many foods because these methods may ruin the consistency and other organoleptic (sensory) qualities of the food. Pasteurisation is a form of microbial control for food that uses heat but does not render the food sterile. Traditional pasteurisation kills pathogens and reduces the number of spoilage-causing microbes while maintaining food quality. The process of pasteurisation was first developed by Louis Pasteur in the 1860s as a method for preventing the spoilage of beer and wine. Today, pasteurisation is most commonly used to kill heat-sensitive pathogens in milk and other food products (eg apple juice and honey) (Figure 7.5). However, because pasteurised food products are not sterile, they will eventually spoil.
The methods used for milk pasteurisation balance the temperature and the length of time of treatment. One method, high-temperature short-time (HTST) pasteurisation, exposes milk to a temperature of 72 °C for 15 seconds, which decreases bacterial numbers while preserving the quality of the milk. An alternative is ultra-high-temperature (UHT) pasteurisation, in which the milk is exposed to a temperature of 138 °C for 2 or more seconds. UHT pasteurised milk can be stored for a long time in sealed containers without being refrigerated; however, the very high temperatures alter the proteins in the milk, causing slight changes in the taste and smell. Still, this method of pasteurisation is advantageous in regions where access to refrigeration is limited.

Refrigeration and Freezing
Just as high temperatures are effective for controlling microbial growth, exposing microbes to low temperatures can also be an easy and effective method of microbial control, with the exception of psychrophiles, which prefer cold temperatures. Refrigerators used in home kitchens or in the laboratory maintain temperatures between 0 °C and 7 °C. This temperature range inhibits microbial metabolism, slowing the growth of microorganisms significantly and helping preserve refrigerated products such as foods or medical supplies. Certain types of laboratory cultures can be preserved by refrigeration for later use.
Freezing below −20 °C may stop microbial growth and even kill susceptible organisms. According to the Australian Department of Agriculture, Fisheries and Forestry the only safe ways that frozen foods can be thawed are in the refrigerator, immersed in cold water changed every 30 minutes or in the microwave, keeping the food at temperatures not conducive for bacterial growth. In addition, halted bacterial growth can restart in thawed foods, so thawed foods should be treated like fresh perishables.
Bacterial cultures and medical specimens requiring long-term storage or transport are often frozen at ultra-low temperatures of −70 °C or lower. These ultra-low temperatures can be achieved by storing specimens on dry ice in an ultra-low freezer or in special liquid nitrogen tanks, which maintain temperatures lower than −196 °C (Figure 7.6).

Pressure
Exposure to high pressure kills many microbes. In the food industry, high-pressure processing (also called pascalisation) is used to kill bacteria, yeast, moulds, parasites and viruses in foods while maintaining food quality and extending shelf life. The application of high pressure between 100 and 800 MPa (sea level atmospheric pressure is about 0.1 MPa) is sufficient to kill vegetative cells by protein denaturation, but endospores may survive these pressures.
In clinical settings, hyperbaric oxygen therapy is sometimes used to treat infections. In this form of therapy, a patient breathes pure oxygen at a pressure higher than normal atmospheric pressure, typically between 1 and 3 atmospheres (atm). This is achieved by placing the patient in a hyperbaric chamber or by supplying the pressurised oxygen through a breathing tube. Hyperbaric oxygen therapy helps increase oxygen saturation in tissues that become hypoxic due to infection and inflammation. This increased oxygen concentration enhances the body’s immune response by increasing the activities of neutrophils and macrophages, white blood cells that fight infections. Increased oxygen levels also contribute to the formation of toxic free radicals that inhibit the growth of oxygen-sensitive or anaerobic bacteria like as Clostridium perfringens, a common cause of gas gangrene. In C. perfringens infections, hyperbaric oxygen therapy can also reduce secretion of a bacterial toxin that causes tissue destruction. Hyperbaric oxygen therapy also seems to enhance the effectiveness of antibiotic treatments. Unfortunately, some rare risks include oxygen toxicity and effects on delicate tissues, such as the eyes, middle ear and lungs, which may be damaged by the increased air pressure.
High pressure processing is not commonly used for disinfection or sterilisation of fomites. Although the application of pressure and steam in an autoclave is effective for killing endospores, it is the high temperature achieved and not the pressure directly that results in endospore death.
Desiccation
Drying, also known as desiccation or dehydration, is a method that has been used for millennia to preserve foods such as raisins, prunes and jerky. It works because all cells, including microbes, require water for their metabolism and survival. Although drying controls microbial growth, it might not kill all microbes or their endospores, which may start to regrow when conditions are more favourable and water content is restored.
In some cases, foods are dried in the sun, relying on evaporation to achieve desiccation. Freeze-drying or lyophilisation is another method of desiccation in which an item is rapidly frozen (“snap-frozen”) and placed under vacuum so that water is lost by sublimation. Lyophilisation combines both exposure to cold temperatures and desiccation, making it quite effective for controlling microbial growth. In addition, lyophilisation causes less damage to an item than conventional desiccation and better preserves the item’s original qualities. Lyophilised items may be stored at room temperature if packaged appropriately to prevent moisture acquisition. Lyophilisation is used for preservation in the food industry and is also used in the laboratory for the long-term storage and transportation of microbial cultures.
The water content of foods and materials that is available to microorganisms (water activity which is different to moisture content), can be lowered by adding solutes such as salts or sugars. The amount of available water in microbial cells is reduced dramatically when in the presence of high concentrations of salts or sugars because water will be drawn from an area of low solute concentration (inside the cell) to an area of high solute concentration (outside the cell) creating a hypertonic environment (Figure 7.7). Many microorganisms do not survive these conditions of high osmotic pressure. Honey, for example, is 80% sucrose, an environment in which very few microorganisms are capable of growing, thereby eliminating the need for refrigeration. Salted meats and fish, like ham and cod respectively, were critically important foods before the age of refrigeration. Fruits were preserved by adding sugar; making jams and jellies. Certain microbes, such as moulds and yeasts, tend to be more tolerant of desiccation and high osmotic pressures and may still contaminate these types of foods.

Radiation
Radiation in various forms, from high-energy radiation to sunlight, can be used to kill microbes or inhibit their growth. Ionising radiation includes X-rays, gamma rays, and high-energy electron beams. Ionising radiation is strong enough to pass into the cell, where it alters molecular structures and damages cell components. For example, ionising radiation introduces double-strand breaks in DNA molecules. This may directly cause DNA mutations to occur, or mutations may be introduced when the cell attempts to repair the DNA damage. As these mutations accumulate, they eventually lead to cell death.
Both X-rays and gamma rays easily penetrate paper and plastic and can therefore be used to sterilise many packaged materials. In the laboratory, ionising radiation is commonly used to sterilise materials that cannot be autoclaved, such as plastic Petri dishes and disposable plastic inoculating loops. For clinical use, ionising radiation is used to sterilise gloves, intravenous tubing, and other latex and plastic items used for patient care. Ionising radiation is also used for the sterilisation of other types of delicate, heat-sensitive materials used clinically, including tissues for transplantation, pharmaceutical drugs and medical equipment.
In Europe, gamma irradiation for food preservation is widely used, although in Australia, gamma irradiation is not accepted as a treatment for foodstuffs other than herbs and spices, herbal teas and some tropical fruits. Packaged dried spices are often gamma-irradiated because the process can penetrate paper, plastic, thin sheets of wood and metal and tissue. Great care must be taken when using X-rays and gamma irradiation. These types of ionising irradiation cannot penetrate thick layers of iron or lead, so these metals are commonly used to protect humans who may be potentially exposed.
Another type of radiation, non-ionising radiation, is commonly used for sterilisation and uses less energy than ionising radiation. It does not penetrate cells or packaging. Ultraviolet (UV) light is one example; it causes thymine dimers to form between adjacent thymines within a single strand of DNA (Figure 7.8). When DNA polymerase encounters the thymine dimer, it does not always incorporate the appropriate complementary nucleotides (two adenines), and this leads to formation of mutations that can ultimately kill microorganisms.
UV light can be used effectively by both consumers and laboratory personnel to control microbial growth on or very close to surfaces. UV lamps are now commonly incorporated into water purification systems for use in homes. In addition, small portable UV lights are commonly used by campers to purify water from natural environments before drinking. Germicidal lamps are also used in surgical suites, biological safety cabinets and transfer hoods, typically emitting UV light at a wavelength of 260 nm. Because UV light does not penetrate surfaces and will not pass through plastics or glass, cells must be exposed directly to the light source.
Sunlight has a very broad spectrum that includes UV and visible light. In some cases, sunlight can be effective against certain bacteria because of both the formation of thymine dimers by UV light and by the production of reactive oxygen products induced in low amounts by exposure to visible light.

Sonication
The use of high-frequency ultrasound waves to disrupt cell structures is called sonication. Application of ultrasound waves causes rapid changes in pressure within the intracellular liquid; this leads to cavitation, the formation of bubbles inside the cell, which can disrupt cell structures and eventually cause the cell to lyse or collapse. Sonication is useful in the laboratory for efficiently lysing cells to release their contents for further research; outside the laboratory, sonication is used for cleaning surgical instruments, lenses and a variety of other objects such as coins, tools and musical instruments.
Filtration
Filtration is a method of physically separating microbes from samples. Air is commonly filtered through high efficiency particulate air (HEPA) filters (Figure 7.9). HEPA filters have effective pore sizes of 0.3 µm, small enough to capture bacterial cells, endospores and many viruses, as air passes through these filters, nearly sterilising the air on the other side of the filter. HEPA filters have a variety of applications and are used widely in clinical settings, in cars and airplanes, and even in the home. For example, they may be found in vacuum cleaners, heating and air-conditioning systems and air purifiers.

Filtration and Hospitals
HEPA filters are also commonly used in hospitals and surgical suites to prevent contamination and the spread of airborne microbes through ventilation systems. HEPA filtration systems may be designed for entire buildings or for individual rooms. For example, burn units, operating rooms or isolation units may require special HEPA-filtration systems to remove opportunistic pathogens from the environment because patients in these rooms are particularly vulnerable to infection.
Membrane Filters
Filtration can also be used to remove microbes from liquid samples using membrane filtration. Membrane filters for liquids function similarly to HEPA filters for air. Typically, membrane filters that are used to remove bacteria have an effective pore size of 0.2 µm, smaller than the average size of a bacterium (1 µm), but filters with smaller pore sizes are available for more specific needs. Membrane filtration is useful for removing bacteria from various types of heat sensitive solutions used in the laboratory such as antibiotic solutions and vitamin solutions. Large volumes of culture media may also be filter sterilised rather than autoclaved to protect heat-sensitive components. Often when filtering small volumes, syringe filters are used, but vacuum filters are typically used for filtering larger volumes (Figure 7.10).

References:
Lackemeyer, M. G., Kok-Mercado, F. D., Wada, J., Bollinger, L., Kindrachuk, J., Wahl-Jensen, V., Kuhn, J. H., & Jahrling, P.B. (2014). ABSL-4 aerobiology biosafety and technology at the NIH/NIAID Integrated Research Facility at Fort Detrick. Viruses, 6(1), 137-150. https://doi.org/10.3390/v6010137.
Large industrial autoclaves that allow for moist-heat sterilisation on a large scale.
Relating to sensory qualities of the food e.g. consistency and taste.
A food-preservation method that uses very high pressure to kill microbes while maintaining food quality, leading to extended shelf life.
A freeze-drying process which removes water from materials under vacuum, avoiding formation of ice and producing water vapour directly to preserve the material.
A process in which frozen water changes directly from ice into water vapor without first becoming a liquid.
The formation of gas bubbles or vapour in a liquid, which can disrupt cell structures and eventually cause the cell to lyse or collapse. This occurs because of sonication – when the use of high-frequency ultrasound waves create rapid pressure changes within a liquid leading to bubble formation and the disruption of cell structures.
Sonication: The use of high-frequency ultrasound waves to create pressure changes in a liquid, causing cavitation (bubble formation and collapse) that disrupts cell structures and can lead to cell lysis.